Technosim Engenharia

3D scanner accuracy: how to read the datasheet correctly

3D scanner accuracy: how to read the datasheet correctly

A 3D scanner's accuracy only makes sense with the test condition next to the number. On the datasheet, accuracy says how far the measurement departs from the true value, resolution says the smallest distinguishable detail and volumetric accuracy shows how the error grows with part size. Confirm everything with your part.

What are accuracy, precision and repeatability in a 3D scanner?

Three words cause most of the confusion. Accuracy (or trueness) is the closeness between the measured value and the true value. Precision, in the metrological sense, is the closeness between repeated measurements of the same thing, that is, the dispersion. Repeatability is that dispersion measured under the same conditions: same operator, same equipment, same part, same day.

A shooting target helps. Shots clustered far from the center are precise and not very accurate. Shots scattered around the center are accurate on average and not very precise. What you want is a tight group at the center. In everyday industrial speech, people say "precision" for everything, and many brochures take advantage of this looseness of language. So, when reading "precision of 0.02 mm," ask which of the three quantities that number represents.

How to read each datasheet field?

The most common fields, and what each one actually says:

  • Accuracy: maximum or typical error under a defined test condition, usually measuring a calibrated artifact in a small volume. It is the most favorable number on the datasheet and rarely represents your day-to-day use.
  • Volumetric accuracy: how the error grows with the measured dimension. It usually comes as a sum, a fixed value plus a value per meter. It is the field that matters for large parts.
  • Resolution: minimum spacing between mesh points. It shows the level of detail the scanner can represent, not how correct the measurement is. Fine resolution with poor accuracy produces a detailed and wrong mesh.
  • Measurement rate: points per second. It indicates capture speed, with no direct relation to error. A high rate shortens field time, and quality comes from optics, calibration and algorithm.
  • Scan area and working distance: define how much surface goes into each capture and at what distance the scanner operates. They affect productivity and the ease of reaching narrow regions.

Why are accuracy and volumetric accuracy different numbers?

Because the assembly error of the captures accumulates. The scanner captures one section at a time and the software joins the sections, whether by adhesive targets, by part geometry or by an optical tracker. Each join adds a small uncertainty, and on larger parts that sum dominates the error.

Data from two Scantech families show the effect. The NimbleTrack states accuracy of up to 0.025 mm and maximum volumetric accuracy of 0.064 mm across the entire range. The KSCAN-Magic states accuracy of up to 0.020 mm and volumetric accuracy of 0.015 mm + 0.012 mm/m when paired with the MSCAN-L15 photogrammetry system. By the formula, on a 2 m part, the volume term gives 0.015 + 0.024 = 0.039 mm. The smaller number on the datasheet applies to the small scale, and the useful number for your 2-meter shaft is in the formula.

Also note that handheld performance without a tracker, with a tracker and with photogrammetry are distinct conditions. The datasheet of a single scanner may show different numbers for each, and all are true in their own condition.

Which test conditions change the number?

A datasheet number only applies under the condition in which it was obtained. The variables that weigh the most:

  • Artifact used in the test: sphere, ball bar or calibrated plane, and its size.
  • Alignment method: with targets, without targets, with photogrammetry or with a tracker.
  • Scanner mode: fast scanning and fine scanning modes usually have different numbers.
  • Temperature and stability of the environment during the test.
  • Whether the result is maximum error or typical value, and how the supplier defines the term.
  • Whether the test followed a published guideline, such as VDI/VDE 2634, or an internal procedure.

A real part adds other factors: surface finish, color, shine, edges, holes and operator skill. That is why the datasheet number is a reference ceiling, not a prediction of your result.

What to ask the supplier before buying?

  1. Which quantity does each number refer to: accuracy, volumetric accuracy, repeatability or resolution?
  2. On which artifact, in which volume and with which alignment method was the test performed?
  3. Did the test follow a published guideline? Is there a test report I can read?
  4. How does the uncertainty grow on a part the size of mine?
  5. How is the periodic verification of the equipment done in Brazil, and with which artifact?
  6. Can I scan my part, with my operator, before deciding?
  7. Who provides support, calibration and training on national soil?

The last question weighs more than it seems. Metrology equipment loses value when it sits idle waiting for service, and the lead time for a repair abroad is only known after it happens.

How to test the scanner with your own part

Testing with the customer's part is worth more than any datasheet. A simple procedure:

  1. Choose a representative part, with the most difficult surfaces of your use: edges, holes, dark or shiny areas.
  2. Beforehand, measure the 5 to 10 features that matter most on an instrument of lower uncertainty (for example, a CMM or a calibrated gauge).
  3. Let the part acclimate in the test environment and note the temperature.
  4. Scan the part three times, with the same setup, to observe repeatability. If possible, have another operator repeat it.
  5. Align each mesh to the reference using the same datum criterion and extract the same features.
  6. Compare the difference to the reference instrument (accuracy in use) and the difference between the three repetitions (repeatability).
  7. Record the total time, including setup, capture, mesh cleaning and reporting.

If the scanner passes on shape features and fails on small holes, that is a useful result: you found out where it works and where it does not. To understand the boundary between the methods, also read the guide on 3D scanner and coordinate measuring machine.

Common mistakes when interpreting accuracy

  • Choosing the scanner by the lowest number on the datasheet, without looking at volumetric accuracy for the size of your part.
  • Confusing resolution with accuracy: a dense mesh does not guarantee a correct measurement.
  • Judging a scanner by its measurement rate, which describes capture speed.
  • Comparing datasheets from manufacturers that used different artifacts, volumes and definitions.
  • Ignoring the part's temperature, especially for aluminum and for parts that have just come out of machining.
  • Evaluating with an easy part, with a matte surface and simple geometry, and then using it on difficult parts.

When the datasheet accuracy is not the main criterion

In many applications, other criteria decide. For reverse engineering of parts without a drawing, the mesh needs to be sufficient for modeling, and an error of tenths of a millimeter is usually acceptable. For 3D printing, the tolerance of the printing process is usually larger than the scanner's error. For hundredth-of-a-millimeter tolerances on holes and planes, the criterion stops being the scanner and becomes another instrument. In that case, digital inspection with a probe or the CMM is the right path.

Frequently asked questions

What is the difference between accuracy and precision in a 3D scanner?

Accuracy is how close the measurement comes to the true value. Precision is how similar repeated measurements are to each other. A scanner can always repeat the same wrong value, being precise and not very accurate. For metrology, you need both, and the datasheet should say which one each number represents.

What is volumetric accuracy?

It is how the measurement error grows with the size of the measured volume. It is usually expressed as a fixed value plus a value per meter. It matters on large parts, because the error accumulated between captures starts to dominate. If your part is more than one meter long, look at this field before accuracy.

Does high resolution mean a more correct measurement?

No. Resolution is the spacing between mesh points and indicates the detail the scanner represents. The correctness of the measurement depends on accuracy and calibration. A very dense mesh with poor accuracy represents a wrong shape in detail. Evaluate both pieces of information together.

How do I verify my scanner's accuracy after purchase?

Periodically scan a reference artifact, such as a calibrated ball bar, and compare the measured distances with the certified ones. Do this before critical jobs and after transport or impact. Record the results to track drift and demonstrate control in audits.

Can I trust the datasheet number?

Trust the condition described alongside the number. It is true in the test that generated it, normally with a calibrated artifact, a controlled environment and a trained operator. To know the performance on your part, request a demonstration with it and compare with an instrument of lower uncertainty.

Do I need periodic calibration of the scanner?

Yes, when the result goes into a quality decision. The frequency and method follow the manufacturer's recommendation and your management system. Between calibrations, perform checks with an artifact. Ask the supplier who performs the service in Brazil and how it is done.

Bring your part to the demonstration

Technosim demonstrates Scantech scanners with the customer's part, and trains teams in the 3D scanner digitization course. To request a quote or demonstration, send the drawing or CAD model, the material, the dimensions, the tolerance of the critical features and the report you need to issue. With that we recommend the equipment and plan the test.

← Back to blog